Method ID: HI8105EN

Moisture Determination in Cooking Oil

Titrator Application Methods: General Methods - r. 0.4

Results:

Titration Report

Method Name:

Moisture in Cooking Oil

Time & Date:

12:00 Jan 1, 2011

Sample Size:

4.0296 g

Std. Titrant Conc.:

1.0000 mg/mL

Drift Value:

3.4 ug/min

End Point Volume:

2.6808 mL

Result:

660 ppm

Titration Duration:

6:30 [mm:ss]

Estimated Cell Volume:

58.11 mL

Titration went to Completion
Operator Name:
Analyst Signature: ________________________