beautypg.com

Auxiliary functions – Hanna Instruments HI 903 User Manual

Page 110

background image

7-4

AUXILIARY FUNCTIONS

7.2

Burette

To access the

Burette

screen, press

from the

Idle

screen.

Highlight the desired option and then press

.

Note: Do not perform burette functions with solvent below the “Min” sign. Doing so could

spray titrant on the beaker top or other components.

7.2.1

Prime Burette

After solvent has been added to the titration cell, the burette can be primed with titrant. The

priming process consists of several cycles of filling and emptying the burette with titrant. It
ensures that any air, water or water vapor in the burette or tubing is removed.
Two rinse cycles of burette are shown in the figure below. The dispensing tube is connected
on the right side and the aspiration tube on the left side.

Note: Before starting this operation, the aspiration tube must be inserted into the titrant

bottle.