Quick start guide – Hanna Instruments HI 903 User Manual

Page 12

background image

12

QUICK START GUIDE

• Highlight the

Setup Titration Report

option and press

.

• Mark the fields to be included with the “*” symbol using the

and

keys and

press

to toggle the selection.

• Press

and then press

to return to the main screen.

Fill Titration Beaker with Solvent

The titration beaker must be filled with 1-component up to the MIN marker (about 50 mL of
solvent):

• From the

Idle

screen, press

.

• Push and hold the FILL button located on the top of the air pump.

• Wait until the beaker is filled up to the MIN marker with solvent.

• Stop the air pump by pressing

and then confirm the approximate amount of

solvent in the beaker.

Prepare the Solvent for Samples

Before beginning a titration, residual moisture inside the titration beaker and solvent must be
reacted:

• From the Idle screen, press

. The titrator will enter Pre-Titration mode, start the

magnetic stirrer, and begin dosing titrant into the titration beaker. If no titrant can be
seen moving through the anti-diffusion tip after several doses, press

and

verify that no titrant is leaking from the burette housing or from the dispensing tube
fittings.

• Once all residual moisture has been reacted (endpoint potential is reached), the

titrator will enter Drift Analysis mode (assuming Automatic Drift Entry is selected).
The titrator calculates the rate of atmospheric moisture seeping into the titration
beaker for the next minute and displays the result in the lower right corner of the
display.

• If the Drift Rate is stable and the endpoint potential is maintained, the titrator will

enter Standby mode. The titrator continues to maintain the endpoint potential and
update the background drift rate.

Preparing and Introducing the Sample

Sample Mass Preparation

Measuring the sample size by mass using an analytical balance will give the most reproducible
results.