Methods – Hanna Instruments HI 903 User Manual

Page 81

background image

5-17

METHODS

Before the titrant standardization is started, the user is asked again to enter the standard size.
The standard size (mass and volume) can be acquired automatically from a compatible
analytical balance (when the balance feature is enabled - see Section

4.11, Setup Balance

Interface

).

5.5.13

Titrant

The user can access the Karl Fischer titrant database and customize related parameters.

Using the

and

keys, highlight the titrant from the existing list and press

to

choose it.

Press

to create a new titrant.

Press

to remove the titrant from the list.

Press

to edit the titrant parameters.