Troubleshooting

72

FIP-400B

Solving Common Problems

An APC fiber is

connected to an

FIP-435B probe, the blue

LED is blinking and the

motor is not running

When the fiber connector is detected, this will then initiate

the auto-focus sequence.

Refresh rate is very low

Ensure that the power saving mode is deactivated.

On your smart device, close the applications that are not

in use.

Reduce the number of probes operating in the vicinity.

Bring the probe and the smart device closer together.

The FIP status LED blinks

red for 2 seconds in Live

Video mode and turns

back to blue (auto focus

timeout)

Capture the image again.

The FIP LED blinks red

for 2 seconds in Capture

mode and no analysis

results are available

There was an analysis error. Repeat the inspection process.

The user interface

displayed in the

application is truncated.

Change font size in the smart device settings to facilitate

reading comprehension.

Problem

Solution