background image

5.4. Digital Linearization Function

Even if the zero and span calibration have been completed, there may still remain some

linearity deviation by the performance of the weighing unit The digital linearization

function can rectify and reduce the linearity deviation using weighing points during the

zero and capacity setting. Up to three weighing points can be specified.

Caution □ This function does not improve repeatability and hysteresis.

□ Use the mass on the condition that inr I < Lnr E < Lnr 3,

□ Do not press the I key while the MD mark is lit

Step 1

Middle point

Rash indication

w

mark

First Point

Lnr 1

Net

Second Point

Lnr 3

Gross

Third Point

i nr 3

Net and Gross

Zero calibration has been finished and Fxxxx

displayed. (Start this procedure from step 10 of

"5.2.7.Span Calibration")

ts

Step 2 Press the SsETTAR^ key to enter this digital

linearization function.

Step 3 The value of the middle point is displayed after

indicating

I

tnr x j. x Is

l,3 or3.

Step 4

Q

Select a way.

Press the iMODEj key to finish this function. Proceed

to step 7 and other points are cleared (canceled).

Select a middle point value using the following keys.

Proceed to step 5.

5]

......... The key to select a figure.

A I key............ The key to select a number.

Step 5 Place a mass equivalent to the displaying value on

the weighing unit. Press the j ^ I key after the MD

mark has turned off. Proceed to step 6.

Step 6 If you add new middle point, proceed step 3, 4, 5.

If you finish this function, proceed to step 7.

Step 7 Perform step 10 of "5.2.7.Span Calibration" on page

16 immediately.

ConHnue after zero calibration.

Span calibration

at step 10.

f

lOOOO

2 seconds

L

jr 1

•>o<*

mo

>

TARE

A

L ¿500

Place the mass

"vZUnlit MD mark

Q

PRIfff

Inr p

*>Tv

ZERO TARE

>

2 seconds

Second point

i SQOO

Place the mass

v^Unlit MD mark

2 seconds

Third point

Q

PRINT

Li

-K>

ZERO

>

TARE

L

1500

Place the mass

0

.

Q

MODE

yiNT

5.4. Digital Lineariz^on Function

Page 22

Return to span catibraion

AD-4329 Weighing Indicator