5.2.2.
Preparation
step 1 Keep the following conditions to calibrate the scale (indicator) correctly.
□ Maintain a constant temperature, stable power and stable input voltage from the
load ceil.
□ Avoid direct sunshine and in the neighborhood of an air conditioner.
□ Do not install the scale (indicator) where there is a strong magnetic field.
Step 2 Turn the display on during 10 minutes.
5.2.3.
Entering the Calibration Mode
Step 3 Press the ICAlj key In the norma! weighing mode. The
two seconds when entering the calibration mode.
CHL
(O is displayed for
MODE! key.... Parameters are not revised and the next item is displayed.
CALI key........Parameters are stored in the indicator and
Snd
I is displayed.
Press the jOperate/Standbvl key to turn the display off.
Press and hold the ¡NET/GROSSl key and press the |CAU key
....... .............All parameters are not changed
and the calibration mode is.
finished. Press the jOperate/Standby
key to turn the display off after
Turn the display on
for 10 miniuies
displaying
CBnCEL
CAL
m
,n
Zero
Calibration^
^
Span Calibration.
5. Calibration
AD-4329 Weighing Indicator