5.2.6.
Zero Calibration
Step 7 Check the
CfiL 0
display.
Select a zero calibration method to adjust the zero point
Weighing input
(Normal way)
The adjustment method
with nothing on the
weighing unit.
To step 8
Digital input
The numerical input to
enter a load cell output
voltage.
To step 9
Weighing Input
Step 8 Place nothing on the weighing unit. Press the
key
after the MD mark has turned off. Proceed to steplO.
MOPg key....... The key not to change the zero point
data and proceed to step 10.
Caution Do not press the [ *-■ key while the MD mark is lit
(detecting motion). Arrange the condition using
the FOO filter function.
Digital Input
Step 9 Pressing the @ key, a stored input voltage parameter
of the zero point is displayed in the unit of mV/V. Adjust
the input voltage using the following keys.
key.....The key to select a figure,
key.....The key to select a number.
key.....The key to store the zero point parameter
and proceed to step 10.
MODE] key..The key not to change parameter and
proceed to step 10.
Zero calibration
To Span Calibration
AD-4329 Weighing Indicator
Page 15
5. Calibration