5.2.7. Span Calibration
Step10 Check the | Fxxxx
I
display. The xxxx is a capacity.
Select a span caliiDration method to adjust the capacity.
Weighing a mass except the
maximum capacity
The method to weigh a mass except
the maximum capacity.
To step 11
Weighing maximum capacity
The method to weigh a mass
equivalent to the maximum capacity.
To step 14
Digital input
The numerical input to enter a load
ceil output voltage.
To step 16
Weighing a Mass except the Maximum Capacity
Step11 Set a mass value using the foilwing keys,
key.....Key to select a figure.
>
key Key to select a number.
Step12 Place a mass equivalent to displayed value on the
weighing unit.
StepIS Press the I ^ j key after the MD mark turns off.
Proceed to step 17.
MOPg key........The key not to change parameter
and go back to step 4.
CAll key...........The key not to change parameter
and proceed to step 17.
Caution Do not press the
J
key white the MD mark Is
tit (detecting motion). Arrange the condition
using the FOQ filter function.
Weighing Capacity Mass
Step14 Place a mass equivalent to the maximum capacity on
weighing unit.
StepIS
Press the ] key after the MD mark turns off.
Proceed to step 17.
MODE! key...... The key not to change parameter
and go back to step 4.
CALI key..........The key not to change parameter
and proceed to step 17.
key while the MD mark is
Caution Do not press the
lit (detecting motion). Arrange the condition
using the FOO filter function.
span calibration
Maximum capacity
Place the displayed
mass on the
weighing unit.
^ Unlit MD mark