Symptom

Possible cause

Action required

Display shows
“Internal
Temperature
P

robe Fault”

Faulty temperature
probe

Have a competent person check the probe for an
open or short circuit fault or contact Grant.

Display shows
“External
Temperature
P

robe Fault”

External temperature
probe not plugged in

Faulty temperature
probe

Check the external probe is properly connected at
the rear panel.

Have a competent person check the probe for an
open or short circuit fault or contact Grant.

Display shows
“Program Fault”

One or more program
segments contains
an invalid setting or
could not be read
from the memory

Reload the program if it came from LabWise or edit
each segment on the unit (TXF200 only)

Display shows
“Overheating-
Power reduced

Heating water at or
near to boiling
without lid


Heating very large
volumes of liquid with
large thermal losses

Add a lid to reduce thermal losses and leave unit
running with the “Overheating” warning present. If
the alarm has not cleared within 1 hour contact
Grant.

Add measures to reduce thermal losses and leave
unit running with the “Overheating” warning present.
If the alarm has not cleared within 1 hour contact
Grant.

Display shows
“Service required
01

Faulty fuse(s), relay
or heater element

Have a competent person check the product or
contact Grant.

Display shows
“Service required
02

Pump or propeller is
obstructed

Faulty motor

Remove obstruction.


Have a competent person check the product or
contact Grant.

Display shows
“Cross-check
Failure

PCB fault

Have a competent person check the product or
contact Grant.