Dial

Important Safety Information

Read the full manual before use. Safety may be impaired if it is not followed.

Only connect to the voltage and frequency shown on the serial number plate. An earth is required.

Always ensure the bath is filled with a liquid rated for the working temperature. See manual for liquid selection.

Ensure the TX150/TXF200 is positioned so that the plug is easily accessible for disconnection.

Before moving or cleaning, allow the unit to cool down and disconnect from the power supply.

Safety can only be assured if the TX150/TXF200 is securely installed on a suitable vessel. See manual for guidance.

Service Department,
Grant Instruments (Cambridge) Ltd
Shepreth, Cambridgeshire, SG8 6GB, UK

T: +44 (0)1763 260 811
F: +44 (0)1763 262 410

E:

[email protected]

W: www.grantinstruments.com

Safety cut out and alarm light

The TX150/TXF200 is fitted with a
float switch which monitors the liquid
level during operation. Should the
liquid level drop below a certain level,
the safety cut out will be activated,
and alarm light will appear. Top up
the bath with liquid until at least a
minimum level to resume operation.

Cleaning

Clean the unit with a damp cloth,
using only water. Do not use
chemical cleaning agents.

Consult manual for full instructions.

Service & repair

For service or repair please contact
our Service Department or your local
distributor. All returned units must be
accompanied by a Return Materials
Authorisation (RMA) number,
obtainable by contacting the Grant
service department.

Using the external pump

Ensure the TX150/TXF200 is switched off.
Connect hoses securely to the pump outlet
plate as shown left.

Always use pump connectors and hoses
that are suitable for the operating
temperature and liquid used.

Never disconnect any pipes or hoses
while they contain very hot or very cold
liquids or while the unit is pumping.

30244 V1 Apr 2012

Optima TX150, TXF200 Quick Start Guide

F (function button)

Over-temperature dial

Full colour screen

S (select button)

Float low liquid level
protection

Pump outlet
plate

Setting the over-temp

Rotate dial to desired

temp cut-out

(above set point)

.

:





.

°c



.

°c

Settings

Stand-by

Probe

Heater

indicator

Preset

Set temp

Timer

Program

Pump

Setting the temperature

:





.

°c



.

°c

Rotate the dial until the
set temperature icon is
highlighted.
Press the S button to
select. Rotate the dial to
the desired temperature.
Press S to set.