11.0 Troubleshooting
Symptom
Possible cause
Action required
Temp continues
to rise when not
expected
Set temp is higher
than liquid temp
Check that the bath set temperature is correct (see
section 5.2.9).
Set temperature
too restricted
Liquid type set does
not allow required set
point
Change to different liquid type (see section 5.3.4).
Temperature
does not rise
when expected
Set temp is lower
than liquid temp
Set temperature is
too close to ambient
Check that the bath set temperature is correct (see
section 5.2.9).
Increase the set temperature (see section 5.2.9) or
fit accessory cooling (see section 4.9).
Display shows
“High
Temperature
Alarm”
High temperature
warning alarm has
tripped
Check that the bath set temperature is correct (see
section 5.2.9).
Check that high temperature alarm is correct
(section 5.3.6).
Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).
Display shows
“Low
Temperature
Alarm”
Low temperature
warning alarm has
tripped
Check that the bath set temperature is correct (see
section 5.2.9).
Check that low temperature alarm is correct (section
5.3.6).
Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).
Unit showing
erratic
temperatures
Calibration values not
set correctly
Restore the factory calibration settings (see section
5.4.1) then re-calibrate if required (see section 5.4).
New calibration
point is not
saved
Incorrect calibration
value
The temperature calibration point is less than 20°C
from an existing calibration point
– choose a higher
temperature (see section 5.4).
The measured liquid temperature is more than 5°C
away from selected calibration temperature.
Only the
Standby icon
can be
highlighted
Unit is in standby
mode
Highlight and Select the Standby icon and Pres
s “S”
to exit Standby mode.
Icons can be
highlighted but
not selected
Program running
Wait until program has finished or stop program.