11.0 Troubleshooting

Symptom

Possible cause

Action required

Temp continues
to rise when not
expected

Set temp is higher
than liquid temp

Check that the bath set temperature is correct (see
section 5.2.9).

Set temperature
too restricted

Liquid type set does
not allow required set
point

Change to different liquid type (see section 5.3.4).

Temperature
does not rise
when expected

Set temp is lower
than liquid temp

Set temperature is
too close to ambient

Check that the bath set temperature is correct (see
section 5.2.9).

Increase the set temperature (see section 5.2.9) or
fit accessory cooling (see section 4.9).

Display shows
“High
Temperature
Alarm”

High temperature
warning alarm has
tripped

Check that the bath set temperature is correct (see
section 5.2.9).

Check that high temperature alarm is correct
(section 5.3.6).

Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).

Display shows
“Low
Temperature
Alarm”

Low temperature
warning alarm has
tripped

Check that the bath set temperature is correct (see
section 5.2.9).

Check that low temperature alarm is correct (section
5.3.6).

Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).

Unit showing
erratic
temperatures

Calibration values not
set correctly

Restore the factory calibration settings (see section
5.4.1) then re-calibrate if required (see section 5.4).

New calibration
point is not
saved

Incorrect calibration
value

The temperature calibration point is less than 20°C
from an existing calibration point

– choose a higher

temperature (see section 5.4).

The measured liquid temperature is more than 5°C
away from selected calibration temperature.

Only the
Standby icon
can be
highlighted

Unit is in standby
mode

Highlight and Select the Standby icon and Pres

s “S”

to exit Standby mode.

Icons can be
highlighted but
not selected

Program running

Wait until program has finished or stop program.

This manual is related to the following products: