5.3.4 Selecting a liquid type


The liquid type determines the limits of the set temperature range.
Selection of the liquid types below changes the settable temperature range as follows:

Liquid

Set temperature range

Water

0°C to 100°C

Water-Glycol

*-30°C to 70°C

Low Temp Oil

-50°C to 50°C

High Temp Oil

70°C to 150°C (TX150)

¥

70°C to 200°C (TXF200)

Override

§-50°C to 150°C (TX150)
§-50°C to 200°C (TXF200)

* Water-Glycol (50% water, 50% antifreeze (inhibited ethylene glycol)

High Temp Oil (silicone fluid with the following characteristics: viscosity 20 centistokes,

flash point ≥230°C, fire point ≥280°C).

¥

High Temp Oil (silicone fluid with the following characteristics: viscosity 50 centistokes,

flash point ≥285°C, fire point ≥340°C).

§

When set to Override it is the user’s responsibility to select a suitable safe liquid.

See section 4.6 for full description of recommended bath liquids.

1. Rotate the dial until the settings icon is

highlighted, press the S button.








2. Rotate the dial to scroll up and down the list until

liquid is highlighted, press S to select.


Liquid options are displayed.


3. Rotate the dial to scroll through the liquid options,

press S to save the selection.


The display returns to the settings screen.