beautypg.com

Running a bath preset, Running a countdown timer – Grant Instruments TXF200 Operating Manual User Manual

Page 18

background image

30424 V6

TX150 & TXF200

Page 17

Operating Manual

www.grantinstruments.com

5.2.11 Running a bath preset


Each TX150/TXF200 contains three presets which can be configured to different set
temperatures and in the case of the TXF200 the pump speed can additionally be
configured. This allows the bath to be conveniently run at frequently used temperatures
and pump speeds. See section 5.3.1 for information on preset configuration.

1. Rotate the dial until the preset icon is

highlighted, press the S button.


2. Rotate the dial to select the desired preset 1, 2

or 3, press S to run the preset.

The preset will automatically start as soon as S
is pressed.

If no key is pressed for 10 seconds or if F is
pressed the preset icon is no longer active and
will remain at its original value.
If the set temperature of the preset is not
allowed for the selected liquid (e.g. a preset of
150°C when using water) then the preset icon
will return to “
“ and the preset will not be used.

Refer to section 4.6 to find the allowable temperatures for each liquid.

5.2.12 Running a countdown timer


The countdown timer on the TX150/TXF200 can be set in the range of 1 minute to 99
hours. The countdown timer will sound a buzzer at the end of a countdown period. See
section 5.3.2 for information on configuring countdown timer expiry actions.

1. Rotate the dial until the countdown timer icon is

highlighted, press the S button.


The countdown timer will display the last
countdown time set.

If no key is pressed for 10 seconds or if F is
pressed the countdown timer setting is no longer
active and will remain at its original value.


2. Rotate the dial to set the desired countdown

time, press S to store.

The countdown timer will begin counting down
from the set time. At the end of the countdown
timer period a buzzer will sound. This can be
cancelled by pressing F or S.

To cancel an active countdown timer:

1. Rotate the dial until the countdown timer icon is highlighted, press the S button.

2. Press the F button to cancel the countdown timer.

The countdown timer is stopped.



.



°c



.

