6.0 Technical specifications


6.1

Operating conditions

Ambient temperature range

5 to 40°C

Altitude above sea level

Up to 2,000m (6,500ft)

Operating environment

Indoor use only

Maximum relative humidity

80% RH up to 31°C decreasing to 50% RH at 40°C


6.2

Electrical details


Mains supply:

220-240V @ 50/60Hz or 110-120V @ 50/60Hz

Pollution degree:

2

Installation category:

II


Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.

6.3

Product performance

TX150

TX150L

TXF200

TXF200L

Settable temperature range

-50°C to 150°C

-50°C to 200°C

Stability (DIN 12876)

±0.01°C

Uniformity (DIN 12876)

±0.05°C

Max pump head pressure

310mBar

530mBar

Max pump flow rate

18l/min

22l/min (adjustable)

Max current consumption

9A

12.5A

9A

12.5A

Heater power

1.8kW

1.4kW

1.8kW

1.4kW


6.4

Bath accessories information

ST5

ST12

ST18

ST26

ST38

Tank capacity (litres)

5

12

18

26

38

Liquid depth min/max (mm)

85/140

85/140

75/130

125/180

125/180

P5

P12

P18

Tank capacity (litres)

5

12

18

Liquid depth min/max (mm)*

85/140

85/140

85/140