Grant Instruments TXF200 Operating Manual User Manual
Page 29

TX150 & TXF200
30424 V6
Operating Manual
Page 28
www.grantinstruments.com
EN
Each alarm can be programmed to latch, activate an audible buzzer and a relay. Alarms
are cancelled by pressing the F or S button or if an alarm condition has been removed.
1. Rotate the dial until the settings icon is highlighted,
press the S button.
2. Rotate the dial to scroll up and down the list until
alarms are highlighted, press S to select.
3. High alarm mode is highlighted, press S to
select.
4. Rotate the dial to select fixed temp, deviation or
disabled. Press S to select.
The high alarm screen is displayed
.
°c
High Alarm mode
Alarm Settings
Low Alarm mode
Save
Fixed Temp
Disabled
Holdoff
:
.
°c
Probe
Settings
Alarms
Buzzer
Language
Liquid temp
Set point
temp
Temp
Time
5
10
15
Hold off time.
Alarm value
0
Hold off period 8 mins
Alarm acknowledged
2
.
°c
.
°c
: