beautypg.com

Configuring high and low temperature alarms – Grant Instruments TXF200 Operating Manual User Manual

Page 28

background image

30424 V6

TX150 & TXF200

Page 27

Operating Manual

www.grantinstruments.com

5.3.6 Configuring high and low temperature alarms


There are three functions in the alarm menu.

1. A high alarm function, which causes an alarm condition when the bath

temperature goes higher than the alarm value. The alarm value can be set as a
fixed temperature above the set temperature or a deviation offset value above the
set temperature. Below is an example of a high temperature deviation alarm,
where the deviation value has been set to 5

˚C.



2. A low alarm function, when the bath temperature goes lower than the alarm

value. The alarm value can be set as a fixed temperature below the set
temperature or a deviation offset value below the set temperature. Below is an
example of a low temperature fixed alarm, where the fixed temperature value has
been set to -20

˚C.

3. A hold-off time can be entered, which is a user adjustable time of between 0 and

21mins for which the alarms remain muted after either button has been pressed
to acknowledge an alarm condition. If the bath returns to the value such that the
alarm level is not exceeded the alarm will cancel. However, if the temperature
remains outside the alarm level the alarm condition will re-occur after this hold-off
time.

Liquid temp

Time

5

10

15



Temp

Alarm
Value -20

˚C

Set temp -10

˚C

Low temperature fixed alarm.

Liquid temp

Set
temperature

Deviation
5°C

Temp

Time

5

10

15

High temperature deviation alarm.





Alarm value

This manual is related to the following products: