Restoring factory calibration settings – Grant Instruments TXF200 Operating Manual User Manual

Page 35

background image

TX150 & TXF200

30424 V6

Operating Manual

Page 34

www.grantinstruments.com

EN

1. Press the F and S buttons simultaneously for 3

seconds.

The calibrate probe menu displays the factory
defined calibration points and the current
temperature of the bath in the top right corner of
the screen
.

2. Rotate the dial to highlight add point, press S to

select.

The calibrate point menu is displayed. If an
external probe has been selected the external
probe icon will be displayed in the top left corner.
Otherwise the settings icon will be displayed when
calibrating the internal probe.

3. Rotate the dial to highlight measured

temperature. Press S to select

4. Rotate the dial until the temperature being shown

on the calibrated thermometer is displayed on
screen. Press S to save the value.


5. Rotate the dial to highlight save, press S to select.

The display returns to the calibrate probe screen.

6. Press F to return to the home screen

Further calibration points can be added by
repeating steps 2 to 5.
Calibration points can be deleted by selecting
delete in the calibrate point screen.

5.4.1 Restoring factory calibration settings


If the thermometer value is entered before the bath temperature is completely stable the
calibration could be poor and liquid temperature readings will be incorrect. If the
TX150/TXF200 is not in accordance with the thermometer following calibration then it
may not have been successful and the unit should be reset, using the restore factory
settings function.

1. Press the F and S buttons simultaneously for 3

seconds.


2. Rotate the dial to highlight restore factory, press

S to select.

Yes/No are displayed.


3. Rotate the dial to toggle to yes, press S to select.

Factory calibration values are restored.

4. Press F to return to the home screen



.



°c

Point 19.86

Calibrate Probe

Point 70.04

Add Point

Restore Factory

Yes



.



°c

Measured Temperature

Calibrate Point



.

Save

Delete



.



°c

Point 19.86

Calibrate Probe

Point 70.04

Add Point

Restore Factory