ALTANA Haze-Gard Dual User Manual
Page 29

28
English
Continuous reading
only in
single mode
OPERATE
Measurement
Place
Sample at
comp-port
Ready:
Compensation
ISO
T
90.8
CONT. MODE
In the normal measuring mode,
individual measurements are started
by pressing the “operate” button or
the foot switch. In Continuous
Mode, the specimen is measured
repeatedly and the result on the
display updated accordingly. This
allows you to run a large specimen
through the device to establish
whether its quality is homogenous.
The Continuous Mode function can
only be implemented for a single
parameter. If you try to switch to
Continuous Mode for the “transm. +
Haze” parameters, the following
warning appears on the display:
Select one parameter and select
Continuous Mode in the SETUP
menu by pressing the “operate”
button.
9.1
Continuous Mode ISO
Transmittance
Transmittance measurement re-
quires a compensation measure-
ment as a reference. A message on
the display will ask for the definition
of the compensation.
Place your specimen at the comp-
port and press the “operate” button.
The “Measurement active” message
will appear on the display and the
device will then start the continuous
measurement. Then place the
specimen at the haze-port. The
values are displayed continuously.
9 Continuous Mode