beautypg.com

ALTANA Micro-Wave-Scan Manual User Manual

Page 58

background image

57

Cleaning the test tile

• Do not use any acetone!

The accuracy of the measurement can be

significantly impacted by using dirty or damaged

standards.
Since the surfaces of the standards are highly

sensitive, cleaning must be undertaken with great

care.
To clean standards, use a new lint-free cloth, dust-

free lens paper or an optical cloth.
Apply only slight pressure as you clean and make

certain there are no large particles stuck in the

cloth that could damage the surface.
For dirt that is difficult to remove, use an optical

cloth dipped in liquid. Then wipe the surface with a

dry optical cloth.
Exact calibration is not possible unless the

standard is in perfect condition. If the condition

of the standard seems doubtful because of its

appearance or measurement errors, we will be

happy to check it for you.

Cleaning and maintenance