beautypg.com

A&D Weighing ML-50 User Manual

Page 69

background image

MS-70 / MX-50 / MF-50 / ML-50

67

16.

16.

16.

16. Specifications

Specifications

Specifications

Specifications

MS-70

MX-50

MF-50

ML-50

Measurement method

400 W halogen lamp, thermogravimetric analysis

Drying temperature
range at sample pan

50°C to 200°C (1°C increments)

Heating pattern

Standard drying, Ramp drying,

Step drying, Quick drying

Standard drying,

Quick drying

Temperature calibration

By Accessory AX-MX-43

---

Sample weight range

0.1 g to 71 g

0.1 g to 51 g

Accuracy: Repeatability of measurement, (Standard deviation)
Moisture

content 1

over 5 g sample

0.01 %

0.02 %

0.05 %

0.1 %

over 1 g sample

0.05 %

0.1 %

0.2 %

0.5 %

Weighing

mode

0.0005

g

0.001

g

0.002

g

0.005

g

Minimum reading

Moisture

content

0.001

%,

0.01

%,

0.1

%

0.01

%,

0.1

%

0.05%,

0.1

%,

1

%

0.1%,

1

%

Weighing

mode

0.0001

g

0.001g

0.002g

0.002g

Measurement programs

Sample weight and termination value is automatically set with
ACCURACY and % display. When drying rate reaches the
termination value, measurement is automatically completed. ( 2)

Standard mode

Standard drying, Ramp drying, Step drying

Standard drying

Sample weight and termination value is automatically set with
ACCURACY and % display. When drying rate reaches the
termination value, measurement is automatically completed. ( 2)

Quick mode

Quick drying

When drying rate is less than preset termination value,
measurement is automatically completed. ( 2)

Automatic mode

Standard drying, Ramp drying, Step drying

Standard drying

After heating for the preset drying time, measurement is
automatically stopped. (1min. to 480 min.)

Timer mode

Standard drying, Ramp drying, Step drying

Standard drying

When pressing the key at any time, measurement is stopped and
the result is decided.

Analys

is

mode

Manual mode

Standard drying, Ramp drying, Step drying

Standard drying

Moisture content (Wet-base)

Moisture content (Dry-base, Atro)

Dry

content

Ratio

Measurement unit

Weight (g)

Number of memory

20 sets

10 sets

5 sets

Data memory function

100 results

50 results

30 results

Communication
function

RS-232C serial interface

Application software for
Windows (CD-ROM)

WinCT-Moisture Analyzing utility

WinCT

Communication

software

---

This manual is related to the following products: