8.1.1.
8.1.1.
8.1.1.
8.1.1. ACCURACY of the Standard Mode and Quick Mode
ACCURACY of the Standard Mode and Quick Mode
ACCURACY of the Standard Mode and Quick Mode
ACCURACY of the Standard Mode and Quick Mode
ACCURACY of measurement can be set either
HI
,
MID.
or
LO
.
The sample quantity is automatically selected by ACCURACY.
The termination value of the analyzing mode is automatically selected by ACCURACY
and minimum scale value of % display.
The analyzing mode is the program to finish the measurement when a change of
moisture content per one minute becomes smaller than a preset termination value.
The settings are as follows: Specify an ACCURACY.
Drying rate: Change of moisture content per one minute [%/min]
Standard Mode
Standard Mode
Standard Mode
Standard Mode
ACCURACY
Model
Minimum
scale
HI
MID.
LO
0.001 %
0.01 %/min
0.02 %/min 0.05 %/min
0.01 %
0.02 %/min
0.05 %/min 0.10 %/min
MS-70
0.1 %
0.10 %/min
0.20 %/min 0.50 %/min
0.01 %
0.02 %/min
0.05 %/min 0.10 %/min
MX-50
0.1 %
0.10 %/min
0.20 %/min
0.05 %
0.05 %/min
0.10 %/min
0.1
%
MF-50
1
%
0.1
%
Preset
Termination
value
ML-50
1
%
0.10 %/min
0.20 %/min
0.50 %/min
Sample quantity
10 g
5 g
1 g
Use Precise
result
↔
Quick
measurement
Quick Mode
Quick Mode
Quick Mode
Quick Mode
ACCURACY
Model
Minimum
scale
HI
MID.
LO
0.001 %
0.02 %/min
0.05 %/min 0.05 %/min
0.01 %
0.05 %/min
0.10 %/min 0.20 %/min
MS-70
0.1 %
0.10 %/min
0.20 %/min 0.50 %/min
0.01 %
0.05 %/min
0.10 %/min 0.20 %/min
MX-50
0.1 %
0.10 %/min
0.20 %/min 0.50 %/min
0.05
%
0.1
%
MF-50
1
%
0.10 %/min
0.20 %/min 0.50 %/min
0.1
%
Preset
Termination
value
ML-50
1
%
0.20 %/min
0.50 %/min 1.00 %/min
Sample quantity
5 g
2 g
1 g
Use Precise
result
↔
Quick
measurement