MS-70 / MX-50 / MF-50 / ML-50

26

8.1.1.

8.1.1.

8.1.1.

8.1.1. ACCURACY of the Standard Mode and Quick Mode

ACCURACY of the Standard Mode and Quick Mode

ACCURACY of the Standard Mode and Quick Mode

ACCURACY of the Standard Mode and Quick Mode

ACCURACY of measurement can be set either

HI

,

MID.

or

LO

.

The sample quantity is automatically selected by ACCURACY.

The termination value of the analyzing mode is automatically selected by ACCURACY

and minimum scale value of % display.

The analyzing mode is the program to finish the measurement when a change of

moisture content per one minute becomes smaller than a preset termination value.

The settings are as follows: Specify an ACCURACY.

Drying rate: Change of moisture content per one minute [%/min]


Standard Mode

Standard Mode

Standard Mode

Standard Mode

ACCURACY

Model

Minimum

scale

HI

MID.

LO

0.001 %

0.01 %/min

0.02 %/min 0.05 %/min

0.01 %

0.02 %/min

0.05 %/min 0.10 %/min

MS-70

0.1 %

0.10 %/min

0.20 %/min 0.50 %/min

0.01 %

0.02 %/min

0.05 %/min 0.10 %/min

MX-50

0.1 %

0.10 %/min

0.20 %/min

0.05 %

0.05 %/min

0.10 %/min

0.1

%

MF-50

1

%

0.1

%

Preset
Termination
value

ML-50

1

%

0.10 %/min

0.20 %/min

0.50 %/min

Sample quantity

10 g

5 g

1 g

Use Precise

result

Quick

measurement


Quick Mode

Quick Mode

Quick Mode

Quick Mode

ACCURACY

Model

Minimum

scale

HI

MID.

LO

0.001 %

0.02 %/min

0.05 %/min 0.05 %/min

0.01 %

0.05 %/min

0.10 %/min 0.20 %/min

MS-70

0.1 %

0.10 %/min

0.20 %/min 0.50 %/min

0.01 %

0.05 %/min

0.10 %/min 0.20 %/min

MX-50

0.1 %

0.10 %/min

0.20 %/min 0.50 %/min

0.05

%

0.1

%

MF-50

1

%

0.10 %/min

0.20 %/min 0.50 %/min

0.1

%

Preset
Termination
value

ML-50

1

%

0.20 %/min

0.50 %/min 1.00 %/min

Sample quantity

5 g

2 g

1 g

Use Precise

result

Quick

measurement