Store Parameters and Finish the Operation
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation
Store Parameters and Finish the Operation
14. Press the
ENTER
key to store the new parameters of
the measurement program to program number 3.
Pressing the key, the weighing mode is automatically
displayed.
When PROG 3 is recalled, the settings can be used.
To cancel the new parameters and return to the
weighing mode, press the
RESET
key.
Timer Mode
Timer Mode
Timer Mode
Timer Mode
Refer to page 32 for Standard Mode, Automatic Mode or Manual Mode
This explanation uses the following parameters and sample displays of MX-50.
Common
Items
Program number.......................................... 4 ( PROG 4 )
Analysis mode ............................................. Timer mode (Symbol: U-t
)
Drying program
Heating pattern ....................Ramp drying (
)
Final drying temperature .....160°C
Ramp time............................5.0 minutes
Drying time...........................10.0 minutes
Measurement
unit ...............................................Moisture
content
Minimum scale value during measurement.. 0.01 %
Minimum scale value of gram display .......... 0.001 g
Sample quantity ........................................... Approximately 5 g
Data memory function.................................. Not used
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
2. Press
the
PROGRAM
key and press the
↓
or
↑
key to select a program number.
3. Press
the
ENTER
key to use the number.
4. The analyzer displays
end
and returns to the
weighing mode.
Caution
If the data memory function is active, the data
memory number (MEM) is displayed in place of the
program number (PROG).
% MOIST
D
/W
/D
PROG 1
0.000
g
PROG "4
end
PROG 4
0.000
g
Program number
Final drying temperature
50°C
Ramp time
Drying time
0.000
g
160°C
Gram display of
weighing mode