A&D Weighing ML-50 User Manual
Page 38
![background image](/manuals/659063/38/background.png)
MS-70 / MX-50 / MF-50 / ML-50
36
8.2.3.
8.2.3.
8.2.3.
8.2.3. Step Drying
Step Drying
Step Drying
Step Drying
Standard Mode, Automatic Mode or
Standard Mode, Automatic Mode or
Standard Mode, Automatic Mode or
Standard Mode, Automatic Mode or Manual
Manual
Manual
Manual M
M
M
Mode
ode
ode
ode
Refer to page 38 for Timer Mode
This explanation uses the following parameters and sample displays of MX-50.
"Temperature 1" can be set higher than "temperature 2" in step drying.
Common
Items
Program number.......................................... 5 ( PROG 5 )
Drying program
Heating pattern ....................Step drying (
)
Drying temperature 1 ..........120°C
Drying temperature 2 ..........160°C
Time......................................5.0 minutes
Measurement
Unit...............................................Moisture
content
Minimum scale value during measurement.. 0.01 %
Minimum scale value of gram display .......... 0.001 g
Sample quantity ........................................... Approximately 5 g
Data memory function.................................. Not used
Items for Standard Mode
Analysis mode ............................................. Standard mode (Symbol: 5td
)
ACCURACY................................................. MID.
Analyzing mode to finish measurement ............Automatic setting by ACCURACY
Items
for
Automatic
Mode
Analysis mode ............................................. Automatic mode (Symbol: U-a
)
Analyzing mode to finish measurement ............0.05 %/min
Items
for
Manual
Mode
Analysis mode ............................................. Manual mode (Symbol: U-m
)
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
2. Press
the
PROGRAM
key and press the
↓
or
↑
key to select a program number.
3. Press
the
ENTER
key to use the number.
4. The analyzer displays
end
and returns to the
weighing mode.
Caution
If the data memory function is active, the data
memory number (MEM) is displayed in place of the
program number (PROG).
% MOIST
D
/W
/D
PROG 1
0.000
g
PROG "5
end
PROG 5
0.000
g
Program number
Time
Drying temperature 2
Drying temperature 1
Stage 1
Stage 2
(ML-50 can not select step drying)