beautypg.com

A&D Weighing ML-50 User Manual

Page 43

background image

MS-70 / MX-50 / MF-50 / ML-50

41

9.

9.

9.

9. Check Function

Check Function

Check Function

Check Function

9.1.

9.1.

9.1.

9.1. Self

Self

Self

Self----Check Function

Check Function

Check Function

Check Function

Use the self-check function to check whether there is an error or inaccurate result.

During the check, the heater is turned on and the temperature sensor is checked.

Caution

Do not put flammable matter near the analyzer.

Do not put anything on the heater cover.


9.1.1.

9.1.1.

9.1.1.

9.1.1. Operation

Operation

Operation

Operation

1. Display the gram unit (of the weighing mode).


2. Press and hold the

PROGRAM

key to

display CH .


3. Put the breeze break ring, pan support, pan

handle and sample pan in order. (Do not put
a sample on the pan.)
Close the heater cover.
Press the

ENTER

to start the check.

If

Clo5e

is displayed, the heater cover is

not closed. When it is closed, the self-check
function is started.



4. The check function needs approximately one

minute.


Good

result ...Displays

CH pa55

, sounds

buzzer and returns to weighing
mode automatically.

Error..............The buzzer sounds and an error

code is displayed.

Refer to 15.5. Error Message for details.

Example:

CH no

error 0

Ht err


0.000

g

CH

CH . . . .

0.000

g

Preparation

CH

pa55

Checking

Do not put sample

Press and hold

Good result

Weighing mode

This manual is related to the following products: