13.2.1. operation operation operation operation, Cal 0 – A&D Weighing ML-50 User Manual

Page 58

background image

MS-70 / MX-50 / MF-50 / ML-50

56

13.2.

13.2.

13.2.

13.2. Calib

Calib

Calib

Calibration of the Weighting Sensor

ration of the Weighting Sensor

ration of the Weighting Sensor

ration of the Weighting Sensor

A standard mass of 20g or 50g can be used for the calibration
A 20g standard mass (AX-MX-41) is recommend.

Caution

Avoid vibration and drafts that affect the calibration. If affected, the analyzer

may be unable to calibrate the weighing sensor.

Use a 20g mass for the calibration, because the height between the weighing

pan and glass-housing is 26 mm. If a tall mass is used, open the glass-housing
and avoid external influence.

13.2.1.

13.2.1.

13.2.1.

13.2.1. Operation

Operation

Operation

Operation

1. Display the gram unit (of the weighing mode).

2. Install the weighing pan, pan support and pan handle.

Close the heater cover.

3. Press and hold the

PROGRAM

key to display

CH

.


4. Press

the

SELECT

key to display

Cal

.


5. Press

the

ENTER

key to display

Cal 0

.

6. If 20g mass is used, ...... Press the

ENTER

key.

Proceed to step 8.

If 50g mass is used, ...... Press the

SELECT

key.

Proceed to step 7.

7. Press

the

or

key to select 50.000g.

Press the

ENTER

key to store it.

end

,

Cal 0

are displayed in order.



8. When

displaying

Cal 0

, press the

ENTER

key

to input "Data to display ZERO". The standard
mass value is displayed (Example: 20g).


9. Open the heater cover and put the standard mass

on the center of the pan and press the

ENTER

key to input "mass value".

end

is displayed.

10. Remove the mass to return to the weighing

mode

If the report for GLP, GMP and ISO is to be

output (Refer to page 60),

glp

is displayed.

The output condition for the report is selected in
the function table.

0.000

g

CH

Cal

Cal

0

50.000

g

end

The input data to display ZERO

The input of mass value

end

glp

20

0.000

g

Standard mass

Preparation

Remove mass

Mass value

Press and hold