A&D Weighing ML-50 User Manual
Page 35
MS-70 / MX-50 / MF-50 / ML-50
33
Select an Analysis Mode
Select an Analysis Mode
Select an Analysis Mode
Select an Analysis Mode
5. Press
the
SELECT
key to select a symbol, it will
blink. Select an analysis mode using the
↓
or
↑
key.
If standard mode is used, select
5td
.
If automatic mode is used, select
U-a
.
If manual mode is used, select
U-m
.
Set ACCURACY for Standard Mode
Set ACCURACY for Standard Mode
Set ACCURACY for Standard Mode
Set ACCURACY for Standard Mode
6. Select
ACCURACY
with
the
SELECT
key.
Select MID. with the
↓
or
↑
key.
When
ACCURACY
is selected, HI, MID. or LO blinks.
Set the Heating Pattern
Set the Heating Pattern
Set the Heating Pattern
Set the Heating Pattern
7. Select drying temperature with the
SELECT
key.
Select ramp drying
of heating pattern with
the
PROGRAM
key.
Set the Final Drying Temperature
Set the Final Drying Temperature
Set the Final Drying Temperature
Set the Final Drying Temperature
8. Set 160°C with the
↓
or
↑
key.
Set the Ramp Time
Set the Ramp Time
Set the Ramp Time
Set the Ramp Time
9. Set 5.0 minutes with the
↓
or
↑
key.
Set the Preset Termination Value for
Set the Preset Termination Value for
Set the Preset Termination Value for
Set the Preset Termination Value for Automatic
Automatic
Automatic
Automatic M
M
M
Mode
ode
ode
ode
10. Select preset termination value with the
SELECT
key. Select 0.05 [%/min] with the
↓
or
↑
key.
Set the
Set the
Set the
Set the Measurement
Measurement
Measurement
Measurement Unit
Unit
Unit
Unit
11. Select measurement unit with the
SELECT
key.
Select a moisture content (based on a wet
sample ) with the
↓
or
↑
key.
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
12. Select % display with the
SELECT
key.
Select 0.01 [%] with the
↓
or
↑
key.
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display
13. Select gram display with the
SELECT
key.
Select 0.001 [g] with the
↓
or
↑
key.
Automatic
mode
0
"%
"M"O"I"S"T
"/"W
"0."0"1
%
"0."0"0"1
g
0.000
g
"1"6"0°C
ACCURACY
MID.
"0."0"5
%/min
0.000
g
5.0 min
0.000
g
""1"0"5°C
Measurement unit
Minimum scale
value of % display
Minimum scale
value of gram
display
Final drying temperature
Ramp time
Ramp drying
ACCURACY
Preset termination value
Standard mode
Manual mode
"U"-"a
0.000
g
Standard
mode
Automatic mode
Manual mode
Analysis mode