beautypg.com

A&D Weighing ML-50 User Manual

Page 27

background image

MS-70 / MX-50 / MF-50 / ML-50

25

8.

8.

8.

8. Measurement

Measurement

Measurement

Measurement Programs

Programs

Programs

Programs

8.1.

8.1.

8.1.

8.1. List of

List of

List of

List of Measurement

Measurement

Measurement

Measurement Programs

Programs

Programs

Programs

There are five analysis modes.

Measurement Programs

Drying Program

Heating pattern

Analyzing mode to Finish

Measurement

MS-70, MX-50,

MF-50

ML-50

Drying

Temperature

Standard

Mode

5td

Standard drying

Ramp drying

Step drying

Standard

drying

Quick Mode

quc

Measurement condition is
automatically selected by
ACCURACY and minimum
value of % display.
When drying rate is less than
preset termination value,
measurement is completed
automatically.

Quick drying

Automatic

Mode

U-a

When drying rate is less than
preset termination value,
measurement is completed
automatically.

Timer Mode

U-t

Sample is dried for a preset
time. 1min. to 480min.

Manual Mode

U-m

Measurement is completed
by the

STOP

key. Max.

480 min.

Standard drying

Ramp drying

Step drying

Standard

drying

50°C

to

200°C

Drying rate: Change of moisture content per one minute [%/min]


This manual is related to the following products: