A&D Weighing ML-50 User Manual
Page 57

MS-70 / MX-50 / MF-50 / ML-50
55
13.
13.
13.
13. Calibration
Calibration
Calibration
Calibration
The moisture content is calculated with a ratio of wet weight and dried weight.
Therefore, the absolute value of weighing does not influence the calculation of the
moisture content, but it is necessary to get precise weighing for GLP, GMP and ISO.
Use a 20g mass or a 50g mass to calibrate the weighing sensor.
When calibrating the weighing sensor, you can output the calibration report adapted
to GLP, GMP and ISO.
There is a certified temperature calibrator (accessory AX-MX-43, only for MS-70 and
MX-50) to calibrate the pan temperature for precise temperature control.
When calibrating the temperature, you can output the calibration report adapted to
GLP, GMP and ISO.
The analyzer can store an ID number to be used in the calibration report. The
number can be used for management and maintenance of the analyzer
13.1.
13.1.
13.1.
13.1. Identification Number (ID No.)
Identification Number (ID No.)
Identification Number (ID No.)
Identification Number (ID No.)
The ID number consists of the following seven characters.
Characters
0
1
2
3
4 5
6
7
8
9
Space
-(hyphen)
Display
0 1 2 3 4 5 6 7 8 9
-
Characters A B C D E F G H
I
J K L M N O P Q
Display
a b C d e f g H i j k l m N o p q
Characters R S T U V W X Y Z
Display
r s t U v w x y z
13.1.1.
13.1.1.
13.1.1.
13.1.1. Setting the ID Number
Setting the ID Number
Setting the ID Number
Setting the ID Number
1. Turn on the analyzer.
The gram unit (of weighing mode) is displayed.
2. Press and hold the
SELECT
key to enter the function
table. Then
Cl adj
is displayed.
3. Press
the
SELECT
key to display
id
.
4. Press
the
ENTER
key.
5. Set the ID number using the following keys.
Example: lab-123
SELECT
key ... Selects a figure.
↓
,
↑
key............. Selects a value for the figure.
ENTER
key..... Stores the ID No. and proceeds to step 6.
6. Press
the
RESET
key to return to the weighing mode.
lab-123
0.000
g
0.000
g
Cl
adj
id
Press and