A&D Weighing ML-50 User Manual

Page 20

background image

MS-70 / MX-50 / MF-50 / ML-50

18

7.

7.

7.

7. Measurement Procedure

Measurement Procedure

Measurement Procedure

Measurement Procedure

7.1.

7.1.

7.1.

7.1. Standard Mode Operation

Standard Mode Operation

Standard Mode Operation

Standard Mode Operation

The standard mode can obtain the moisture content with settings of ACCURACY

and drying program (heating pattern, drying temperature).

7.1.1.

7.1.1.

7.1.1.

7.1.1. ACCURACY

ACCURACY

ACCURACY

ACCURACY

ACCURACY of measurement can be set either HI, MID. or LO.

The sample quantity is automatically selected by ACCURACY.

The termination value of the analyzing mode is automatically selected by

ACCURACY and minimum scale value of % display.

The analysis mode is the program to finish the measurement when a change of

moisture content per one minute becomes smaller than a preset termination value.

The settings are as follows: Specify an ACCURACY.

ACCURACY

Model

Minimum

scale

HI

MID.

LO

0.001 %

0.01 %/min

0.02 %/min 0.05 %/min

0.01 %

0.02 %/min

0.05 %/min 0.10 %/min

MS-70

0.1 %

0.10 %/min

0.20 %/min 0.50 %/min

0.01 %

0.02 %/min

0.05 %/min 0.10 %/min

MX-50

0.1 %

0.10 %/min

0.20 %/min

0.05 %

0.05 %/min

0.10 %/min

0.1

%

MF-50

1

%

0.1

%

Preset
Termination
value

ML-50

1

%

0.10 %/min

0.20 %/min

0.50 %/min

Sample quantity

10 g

5 g

1 g

Use Precise

result

Quick

measurement

7.1.2.

7.1.2.

7.1.2.

7.1.2. Operation

Operation

Operation

Operation

This operation explanation uses the following example of the MX-50:

Refer to "8.Selection of Measurement Method" for detail.


Input

Parameters

Analysis mode ................................... Standard mode

Drying temperature ............................ 130 °C

ACCURACY....................................... LO

Sample quantity ................................. Approximately 1 g (Automatic selection)

Analysis mode ................................... 0.10 %/min (Automatic selection)


Stored Parameters (Factory Settings for the MX-50)

Heating pattern .................................. Standard drying (

)

Measurement unit ............................... Moisture content is based

on a wet sample

Minimum scale value of % display........ 0.01 %

Minimum scale value of gram display ... 0.001 g

Data memory function........................ Not used

% MOIST

D

/W

/D