A&D Weighing ML-50 User Manual

Page 23

background image

MS-70 / MX-50 / MF-50 / ML-50

21

7.2.

7.2.

7.2.

7.2. Quick Mode Operation

Quick Mode Operation

Quick Mode Operation

Quick Mode Operation

The quick mode can obtain the moisture content with

settings of ACCURACY and drying temperature.

Sample is heated up at 200°C for approximately three

minutes so that moisture content is measured quickly.

7.2.1.

7.2.1.

7.2.1.

7.2.1. ACCURACY

ACCURACY

ACCURACY

ACCURACY

ACCURACY of measurement can be set either HI, MID. or LO.

The sample quantity is automatically selected by ACCURACY.

The termination value of the analysis mode is automatically selected by

ACCURACY and minimum scale value of % display.

Analysis mode is the program to finish the measurement when a change of moisture

content per one minute becomes smaller than a preset termination value.

The settings are as follows: Specify an ACCURACY.

ACCURACY

Model

Minimum

scale

HI

MID.

LO

0.001 %

0.02 %/min

0.05 %/min 0.05 %/min

0.01 %

0.05 %/min

0.10 %/min 0.20 %/min

MS-70

0.1 %

0.10 %/min

0.20 %/min 0.50 %/min

0.01 %

0.05 %/min

0.10 %/min 0.20 %/min

MX-50

0.1 %

0.10 %/min

0.20 %/min 0.50 %/min

0.05

%

0.1

%

MF-50

1

%

0.10 %/min

0.20 %/min 0.50 %/min

0.1

%

Preset
Termination
value

ML-50

1

%

0.20 %/min

0.50 %/min 1.00 %/min

Sample quantity

5 g

2 g

1 g

Use Precise

result

Quick

measurement


7.2.2.

7.2.2.

7.2.2.

7.2.2. Operation

Operation

Operation

Operation

This operation explanation uses the following example of the MX-50 :

Refer to "8.Selection of Measurement Method" for detail.


Input

Parameters

Analysis mode ................................... Quick mode

Heating pattern .................................. Quick drying (

, Automatic selection)

Drying temperature ............................ 130 °C

ACCURACY....................................... LO

Sample quantity ................................. Approximately 1 g (Automatic selection)

Analyzing mode ................................. 0.20 %/min (Automatic selection)


Stored Parameters (Factory Settings for the MX-50)

Measurement unit ............................... Moisture content is based

on a wet sample

Minimum scale value of % display........ 0.01 %

Minimum scale value of gram display ... 0.001 g

Data memory function........................ Not used

% MOIST

D

/W

/D

200°C

Approx. 3 min.

Drying temperature

Heating pattern