Appendix e - calibration procedures – Brookfield DV-II+ Programmable Viscometer User Manual
Page 54
Brookfield Engineering Labs., Inc.
Page 54
Manual No. M/97-164-F1102
Appendix E - Calibration Procedures
The accuracy of the DV-II+ is verified using viscosity standard fluids which are available from
Brookfield Engineering Laboratories or your local Brookfield agent. Viscosity standards are Newtonian,
and therefore, have the same viscosity regardless of spindle speed (or shear rate). Viscosity standards,
calibrated at 25
°
C, are shown in Table E1 (Silicone Oils) and Table E2 (Mineral Oils).
Container size:
For Viscosity Standards < 30,000 cP, use a 600 ml Low Form Griffin Beaker
having a working volume of 500 ml.
For Viscosity Standards
≥
30,000 cP, use the fluid container.
Inside Diameter: 3.25"(8.25cm)
Height:
4.75"(12.1cm)
Note: Container may be larger, but may not be smaller.
Temperature:
As stated on the fluid standard label: (
+
/-) 0.1
°
C
Conditions:
The DV-I
+
should be set according to the operating instructions. The water bath
must be stabilized at test temperature. Viscometers with the letters “LV” or
“RV” in the model designation must have the guard leg attached (see page 34 for
more information on the guard leg).
Normal 25
°
C Standard Fluids
High Temperature Standard Fluids
Viscosity (cP)
Viscosity (cP)
Three Viscosity/Temperatures**
5
5,000
HT-30,000
10
12,500
HT-60,000
50
30,000
HT-100,000
100
60,000
500
100,000
**25
°
C, 93.3
°
C, 149
°
C
1,000
Refer to Brookfield catalog for
more information.
Table E1
MINERAL OIL VISCOSITY STANDARD FLUIDS
BEL Part No.
Viscosity (cP) 25
°
C
B31
31
B210
210
B750
750
B1400
1,400
B2000
2,000
B11000
11,000
B20000
20,000
B80000
80,000
B200000
200,000
B420000
420,000
Table E2