Page 39

Manual No. M/97-164-F1 102

Click the Delete button to delete the B.E.V.I.S. command in the selected row of the
program grid.

Click the Up button to move the B.E.V.I.S. command in the selected row of the
program grid up one row.

Click the Down button to move the B.E.V.I.S. command in the selected row of the
program grid down one row.

Click the Clear button to clear the grid of all B.E.V.I.S. commands. Once cleared, the
commands cannot be retrieved.

Downloads a B.E.V.I.S. program from the PC to the DV-II+

Exits the DVLOADER software program and returns the user to the WINDOWS
Program Manager.

IV.4 Downloading a B.E.V.I.S. Program

Before down loading a B.E.V.I.S. program to the viscometer, ensure the following have been
checked:

The appropriate cable (BEL Part# DVP-80) is connected between the selected COM port of
the PC and the viscometer.

The DV-II+ motor must be OFF.

Set PC PROG to “OFF.”

The Programmable DV-II+ is at the down load screen: OPTIONS|DOWNLOAD A
PROG|LOAD TO SLOT#x
where x is slot 1,2,3, or 4. See Section III.6.

With the LOAD TO SLOT#x screen displayed, choose a storage slot using the DV-II+
arrow keys then press the ENTER/AUTO RANGE key on the viscometer. If after 5
seconds, the viscometer cannot communicate with the DVLoader program, the B.E.V.I.S.
ERROR NO PC ATTACHED
message is displayed and a beeping sound is heard. If a
connection is established, the Down Load button on the PC software becomes enabled , and
the DV-II+ screen displays DOWNLOAD PROG TO STORAGE SLOT #1. Click on
this button to down load the displayed program to the DV-II+. When the down load is
complete, the DV-II+ displays DOWNLOAD DONE TO EXIT PRESS A KEY.

At this point the program in the DV-II+ can be printed and/or run from the viscometer.

Click on this button to exit the DVLOADER software.

Note:

This is an important final step because it allows the RS232 port to be used for