beautypg.com

Appendix e - calibration procedures – Brookfield DV-II+ Programmable Viscometer User Manual

Page 52

background image

Brookfield Engineering Labs., Inc.

Page 52

Manual No. M/97-164-F1102

SC4-14

14

125

0.4

SC4-15

15

50

0.48

SC4-16

16

128

0.2929

SC4-18

18

3.2

1.32

SC4-21

21

5

0.93

SC4-25

25

512

0.22

SC4-27

27

25

0.34

SC4-28

28

50

0.28

SC4-29

29

100

0.25

SC4-31

31

32

0.34

SC4-34

34

64

0.28

SC4-37

37

25

0.36

CPE-40

40

0.327

7.5

CPE-41

41

1.228

2

CPE-42

42

0.64

3.8

CPE-51

51

5.12

3.84

CPE-52

52

9.83

2

V-71

71

2.62

NA

V-72

72

11.10

NA

V-73

73

53.50

NA

Table D2 lists the model codes and spring torque constants for each Viscometer model.

Table D1 (continued)

Table D2

LVDV-II+

0.09373

LV

2.5LVDV-II+

0.2343

2.5 LV

5LVDV-II+

0.4686

5 LV

1/4 RVDV-II+

0.25

1/4 RV

1/2 RVDV-II+

0.5

1/2 RV

RVDV-II+

1

RV

HADV-II+

2

HA

2HADV-II+

4

2 HA

2.5HADV-II+

5

2.5 HA

HBDV-II+

8

HB

2HBDV-II+

16

2 HB

2.5HBDV-II+

20

2.5 HB

5HBDV-II+

40

5 HB

MODEL

TK

MODEL CODE

ON DV-II+ SCREEN