Appendix e - calibration procedures – Brookfield DV-II+ Programmable Viscometer User Manual
Page 52
Brookfield Engineering Labs., Inc.
Page 52
Manual No. M/97-164-F1102
SC4-14
14
125
0.4
SC4-15
15
50
0.48
SC4-16
16
128
0.2929
SC4-18
18
3.2
1.32
SC4-21
21
5
0.93
SC4-25
25
512
0.22
SC4-27
27
25
0.34
SC4-28
28
50
0.28
SC4-29
29
100
0.25
SC4-31
31
32
0.34
SC4-34
34
64
0.28
SC4-37
37
25
0.36
CPE-40
40
0.327
7.5
CPE-41
41
1.228
2
CPE-42
42
0.64
3.8
CPE-51
51
5.12
3.84
CPE-52
52
9.83
2
V-71
71
2.62
NA
V-72
72
11.10
NA
V-73
73
53.50
NA
Table D2 lists the model codes and spring torque constants for each Viscometer model.
Table D1 (continued)
Table D2
LVDV-II+
0.09373
LV
2.5LVDV-II+
0.2343
2.5 LV
5LVDV-II+
0.4686
5 LV
1/4 RVDV-II+
0.25
1/4 RV
1/2 RVDV-II+
0.5
1/2 RV
RVDV-II+
1
RV
HADV-II+
2
HA
2HADV-II+
4
2 HA
2.5HADV-II+
5
2.5 HA
HBDV-II+
8
HB
2HBDV-II+
16
2 HB
2.5HBDV-II+
20
2.5 HB
5HBDV-II+
40
5 HB
MODEL
TK
MODEL CODE
ON DV-II+ SCREEN