Page 11

Manual No. M/97-164-F1 102

II.2

SELECT

SPINDLE

SELECT

SPINDLE

Spindle Selection

LVDV-II+ Viscometers are provided with a set of four spindles and a narrow guardleg; RVDV-II+
Viscometers come with a set of six spindles and a wider guardleg; HADV-II+ and HBDV-II+
Viscometers come with a set of six spindles and no guardleg. (See Appendix E for more information
on the guardleg.)

The spindles are attached to the viscometer by screwing them onto the lower shaft. Note that the
spindles have a left-hand thread. The lower shaft should be secured and slightly lifted with one hand
while screwing the spindle to the left. The face of the spindle nut and the matching surface on the
lower shaft should be smooth and clean to prevent eccentric rotation of the spindle. Spindles can be
identified by the number on the side of the spindle coupling nut.

The DV-II

+

must have a Spindle Entry Code number to calculate Viscosity, Shear Rate and Shear

Stress values. The DV-II

+

memory contains parameters for all standard Brookfield spindles

including custom spindles and the two digit entry code for each spindle (the complete list of entry
codes may be found in Appendix D).

Note:

The DV-II+ will remember the Spindle Entry Code which was in use when the
power was turned off.

Pressing the

SELECT SPINDLE

key will display the current selected spindle code instead of

temperature and cause the character S to begin to blink . It will blink for about three seconds. If the

UP

or

DOWN

ARROW

keys are pressed (while S is blinking) the two character spindle value to the right

of the S character will begin to change (in either an increasing or decreasing direction depending upon
which

ARROW

key is pressed) for each press of the key. If the

ARROW

key is pressed and held, the

display will scroll through the spindle codes for as long as the

ARROW

key is depressed. When it

reaches the last item in the list (either at the top or bottom of the list) the spindle code displayed will
“roll-over” to either the first or last spindle code and the scroll action will continue.

When the desired spindle code is displayed, release the

ARROW

key to halt further scrolling. Press

the

SELECT SPINDLE

key once again. This will cause the S character to cease blinking and the new

spindle code will be accepted for use in viscometer calculations.

Note: You have approximately three seconds in which to press the SELECT SPINDLE

key before the

blinking stops. If you fail to press the SELECT SPINDLE key

before the

blinking stops you will have to repeat the above steps and re-select

the desired spindle.

The DV-II

+

will begin to calculate using the new spindle parameters as soon as the

SELECT SPINDLE

key is pressed the second time.

Note: The number 99 spindle is for use with special spindles when using Brookfield’s

WINGATHER computer program. Refer to the WINGATHER operator manual for
further information on using
“99” spindles.

The DV-II

+

may also be programmed at Brookfield Engineering for “special” user spindles. These

“special” spindles will appear on the spindle scroll list starting with designation “AA” and continuing
through “AZ”. Contact Brookfield Engineering regarding your needs for special spindles.