beautypg.com

Appendix a - cone/plate viscometer set-up, Iv. 5, Example programs – Brookfield DV-II+ Programmable Viscometer User Manual

Page 40

background image

Brookfield Engineering Labs., Inc.

Page 40

Manual No. M/97-164-F1102

Program 2:

For use with an external temperature controller

Command

Command Description

Parameter

Comments

WTP

Wait for temperature

40.0

wait until temperature = 40

°

C (as

an example, a
Thermosel/Controller can be used
for temperature control)

WTI

Wait for time interval

05:00

soak time; allow temperature to

settle

SSN

Set viscometer speed

25.0

run at 25 RPM

SPI

Set print interval

00:30

begin printing data at 30 second

intervals

WTI

Wait for time interval

06:00

wait at this step for 6 minutes,
effectively printing 12 data lines

Program 3:

Wait for cure

Command

Command Description

Parameter

Comments

SSP

Set spindle

31

set to a number 31 spindle

SSN

Set viscomter speed

100.0

run at 100 RPM

SPI

Set print interval

00:05

begin printing data at 5 second

intervals

WPT

Wait for % torque

85.0

wait until % torque = 85; a curing
cycle

Program 4:

Spring relax

Command

Command Description

Parameter

Comments

WKY

Wait for a keypress

Wind to 100%

tell operator to wind spindle until

100% torque is reached

WPT

Wait for % torque

100.0

wait until 100% torque is reached

WKY

Wait for a keypress

Release spindle

tell operator to release the spindle

SPI

Set print interval

00:01

begin printing data at 1 second

intervals

WPT

Wait for % torque

0.0

wait for spindle to completely
unwind to 0% torque

Program 1:

Pre-shear

Command

Command Description

Parameter

Comments

PLN

Print text now

Preshearing now

print user message

SSN

Set viscometer speed

50.0

run at 50 RPM

WPT

Wait for % torque

90.0

wait unitl 90% torque is reached

PLN

Print text now

Collecting data

print user message

SPI

Set print interval

00:10

begin printing data at 10 second

intervals

SSN

Set viscometer speed

10.0

run at 10 RPM

WTI

Wait for time interval

01:40

wait at this step for 1 minute and

40 seconds, effectively printing
10 data lines

IV. 5 Example Programs

The following example programs can also be found on the DVLoader disk that was included with
the DV-II+ Programmable Viscometer:

(time to torque measurement)