I.3
Specifications
Speeds:
Interleaved: LV/RV (18 speeds)
Sequential: LV/RV (18 speeds)
Custom: 54 speeds, user selectable
Note: Refer to Appendix F for detailed list of all speeds.
Weight:
Gross Weight
23
lbs.
10.5 kg.
Net Weight
20
lbs.
9 kg.
Carton Volume
1.65 cu. ft.
0.05 m
3
Temperature sensing range:
-100˚C to 300˚C (-148˚F to 572˚F)
Analog Torque Output:
0 - 1 Volt DC (0 - 100% Torque)
Analog Temperature Output:
0 - 375 Volts DC (-100
°
C to +275
°
C)
RS232 Compatible Serial Port for use with an attached printer or PC.
Centronics Compatible Parallel Port for use with an attached printer.
Viscosity Accuracy:
±
1.0% of full scale range
Viscosity Repeatability:
±
0.2%
Temperature Accuracy:
±
1
°
C : -100
°
C to +149
°
C
±
2
°
C : +150
°
C to +300
°
C
Electrical Certifications:
CUL, CE
Operating Environment:
0
°
C to 40
°
C temperature range (32
°
F to 104
°
F)
20% - 80%R.H.: non-condensing atmosphere
Electrical Certifications:
Conforms to CE Standards:
BSEN 50081-1:
Emission Standard - Light Industrial
BSEN 50082-1:
Immunity Standard - Light Industrial
BSEN 50081-2:
Emission Standard - Industrial
BSEN 50082-2:
Immunity Standard - Industrial
BSEN 61010-1:
Safety requirements for electrical equipment, for measurement, control
and laboratory use
Approved Standards: CSA Class 8721 84 - Electrical Equipment for Laboratory Use
This product has been certified to the applicable CSA and ANSI/UL Standards, for use in
Canada and the U.S.
Installation Category (over voltage category) II: Classification of parts of installation
systems or circuits in local level, portable equipment, appliances, etc. (based on table 1 of
IEC 664).