Specifications

Speeds:

Interleaved: LV/RV (18 speeds)
Sequential: LV/RV (18 speeds)
Custom: 54 speeds, user selectable

Note: Refer to Appendix F for detailed list of all speeds.

Weight:

Gross Weight

23

lbs.

10.5 kg.

Net Weight

20

lbs.

9 kg.

Carton Volume

1.65 cu. ft.

0.05 m

3

Temperature sensing range:

-100˚C to 300˚C (-148˚F to 572˚F)

Analog Torque Output:

0 - 1 Volt DC (0 - 100% Torque)

Analog Temperature Output:

0 - 375 Volts DC (-100

°

C to +275

°

C)

RS232 Compatible Serial Port for use with an attached printer or PC.

Centronics Compatible Parallel Port for use with an attached printer.

Viscosity Accuracy:

±

1.0% of full scale range

Viscosity Repeatability:

±

0.2%

Temperature Accuracy:

±

1

°

C : -100

°

C to +149

°

C

±

2

°

C : +150

°

C to +300

°

C

Electrical Certifications:

CUL, CE

Operating Environment:

0

°

C to 40

°

C temperature range (32

°

F to 104

°

F)

20% - 80%R.H.: non-condensing atmosphere

Electrical Certifications:

Conforms to CE Standards:
BSEN 50081-1:

Emission Standard - Light Industrial

BSEN 50082-1:

Immunity Standard - Light Industrial

BSEN 50081-2:

Emission Standard - Industrial

BSEN 50082-2:

Immunity Standard - Industrial

BSEN 61010-1:

Safety requirements for electrical equipment, for measurement, control
and laboratory use

Approved Standards: CSA Class 8721 84 - Electrical Equipment for Laboratory Use

This product has been certified to the applicable CSA and ANSI/UL Standards, for use in
Canada and the U.S.

Installation Category (over voltage category) II: Classification of parts of installation
systems or circuits in local level, portable equipment, appliances, etc. (based on table 1 of
IEC 664).