B-5
Error
Code
Error Cause
data can be archived.
88
This error results when an attempt is made to archive data during sampling.
Archive operations must occur only when the unit is idle.
97
This error indicates an attempt to use a channel that does not exist on
LabPro (for example, channel 42).
98
This error indicates an undefined error has occurred.
99
This error indicates that the current load on the analog or digital ports is
more than can be supplied by the unit and the power has been turned off to
prevent damage. Do not attempt to restart sampling until the problem has
been corrected.
970
Out Of Data Memory. This is when you collect too much data. Usually there
is another error that happens if you try to setup 4 channels and 12K data
points but if you setup an analog channel and a digital channel, the overflow
depends on how many events you detect. This error happens when the data
pointers overlap.