beautypg.com

Command 9 request channel data – Vernier LabPro User Manual

Page 48

background image

Revision Date: 08/02/02

LabPro Technical Manual

48

Command 9

Request Channel Data

This command requests one data point before sampling starts. Used to verify that setup is correct.

Syntax: {9,channel, mode}

Parameter List:
channel

– indicates the channel on which this command operates. The possible values are

Channel number

Description

1 – 4

Analog channels 1 through 4

11, 12

Sonic Channel 1, Sonic Channel 2

mode

– chooses between present setup or setup for previously collected run. This allows sensors to
be disconnected after run is completed but still able to retrieve auto ID value.

Reqtype value

Description

0

Re-tests the channel for AutoID value

1

returns the stored AutoID value

Return values: one value is returned – current sensor value (curval). Calculators must send a
Get to retrieve these values.

curval

= immediate value of sensor which was setup in requested channel.

Example 1: Retest the analog channel 1 for a smart sensor. In the case of the calculator, store the data in
list L1.
Computer

Calculator

s{9,1,0}

:Send({9,1,0})
:Get(L1)

In a terminal session, the host-LabPro conversation appears as follows:
Host

LabPro

s{9,1,0}

{ +1.94383E+00 }

s{9,1,0}

{ +1.70452E+00 }

s{9,1,1}

{ +1.84005E+00 }

** Change to Barometer sensor **
s{9,1,0}

{ +2.19536E+00 }

s{9,1,0}

{ +2.19536E+00 }

s{9,1,0}

{ +2.19536E+00 }