beautypg.com

Command 102 port power control command – Vernier LabPro User Manual

Page 56

background image

Revision Date: 08/02/02

LabPro Technical Manual

56

Command 102

Port Power Control Command

This command controls how the LabPro powers the analog and DIG/SONIC ports.

Syntax: {102,pwrctl}

Parameter List:
pwrctl

– determines how power is used in LabPro

pwrctl value

Description

-2

Always ON; will automatically power down if running on batteries.
Leaves power to the sensor ports on at all times.

This is valid for LabPro version 6.06210 and newer.

0

Normal mode; “power save mode”

1 - 999

Specifies when power comes on relative to the time a data point is
to be taken. This number specifies the number of seconds before
the reading when power is turned on.

Return values: No information is returned.

Example 1: Turn the LabPro on and leave it on as you might before controlling the digital lines or
outputting a voltage.
Computer

Calculator

s{102,-2}

:Send({102,-2})