Labpro hardware, Connector pinouts – Vernier LabPro User Manual
Page 73

Revision Date: 08/02/02
LabPro Technical Manual
73
LabPro Hardware
Connector Pinouts
LabPro sensors use 6-pin British Telecom-style connectors. Pin-1 is always farthest from the tab.
Pin
Analog CH1, CH2, CH3
Analog CH4
DIG/SONIC1 & DIG/SONIC2
(Right-hand Connector)
(Right-hand Connector)
(Left-hand Connector)
Sonic Digital
1
Vin (-10 to +10 V)
Analog Out and/or
Vin (+/- 5V only)
Echo
D0
2
Gnd
Gnd
Init
D1
3
Vres/Smart ID Data
Vres/Smart ID Data
Auto-ID
D2
4
Auto-ID/ Smart ID CLK
Auto-ID/ Smart ID CLK
+5 Volts DC
+5 Volts
DC
5
+5 Volts DC
+5 Volts DC
Gnd
Gnd
6
Vin-low (0 to 5 V)
Vin-low (0 to 5 V)
Not Applicable
D3
Vin
Vin-low
Channels:
CH1, CH 2, CH 3, CH 4
CH1, CH2, CH3, CH 4
Input signal:
Analog data
Analog data
Input range:
±10 Volts (CH1–3) ±5 Volts
(CH4)
0 to 5 Volts
Resolution (using LabPro’s
bit
−
12
A/D converter):
4.9 mV
1.2 mV
Input impedance:
1.0 M
Ω
>10 M
Ω
•
Vres:
Output reference voltage from LabPro through a 15 Kohm resistor. When using this feature,
Vres should be tied to Vin-low and the value to be measured should be connected between Vin-
low and Gnd.
•
Gnd:
Ground (common for all channels).
•
Auto-ID:
Auto-ID
sensor detection data input. (
Auto-ID
resistor connected from pin 4 to ground.)
•
Echo:
Ultrasonic motion detector input.
•
Init:
Distance initialization signal
•
D0 In/Out
to
D3 In/Out:
Input or output pins for digital pulses.
•
Smart ID Clk: Clock to synchronize data transfer from smart probes.