Reseting calibration – Rice Lake MSI-9750A CellScale RF Portable Indicator User Manual
Page 73

MSI CellScale
®
System • 9750A User Guide Page 73
MEASUREMENT SYSTEMS INTERNATIONAL
Firmware Version 5-XX for 2450 Modems
ENTER CAPACITY
ESC exits no change
ENTER saves value
0-9 replaces value
17
min = 2, max =999999
^ incs, v decs value
CHANNEL / CALIBRATION
--Channel Name--
--Chn-- --Index--
1 Channel is ACTIVE
2 Filter LOW
3 Calibrate Settings
4 Start Calibration
5 Zero RUSure Msg ON
ENTER to
Recalibrate
CLR to Reset Cal
ESC to Exit Cal
CALIBRATION UNITS
1 Pounds
2 Kilograms
3 Tons (US Short)
4 Metric Tons
5 DekaNewtons
6 Ounces
7 Grams
next Channel
available
SCALE DIVISIONS (d)
-Pounds-
1 1 7 0.1
2 2
8 0.05
3 5
9 0.02
4 10
0 0.01
5 0.5
F3 0.005
6 0.2
ZERO CALIBRATION
Unload scale. Press
ENTER when stable.
ESCape ends Cal.
SPAN CALIBRATION 1
Scale =
-XXXXXXX-
Cal Wt 1 =
-XXXXXXX-
0-9 New Cal Weight
When loaded & stable
press ENTER to Cal
ESCape to end Cal
SPAN CALIBRATION
-2-
Scale =
-XXXXXXX-
Cal Wt 2 =
-XXXXXXX-
0-9 New Cal Weight
When loaded & stable
press ENTER to Cal
ESCape to end Cal
RESETING
CALIBRATION
Up to 9 Span Points
Shortcut
4 CalibrationDisabled
CALIBRATION
DISABLED, CYCLE
CELLSCALE POWER
TO CONTINUE
OR
CALIBRATION UNITS
CHANNEL / CALIBRATION
--Channel Name--
--Chn-- --Index--
CALIBRATION LOCK
PRESS SEALSWITCH
ON CELLSCALE TO
CONTINUE
If Disabled (Sealed)
ESC
!
SETUP
F2
CHANNEL
1
ABC
7
STU
F2
CHANNEL
ZERO
CLR
0
ENTER
4
JKL
10) “SPAN CALIBRATION 1” – The CellScale transmits back the
capacity as a possible calibration weight (Cal Wt 1 =). Load the
scale with the fi rst calibration weight. If your cal weight is not equal
to capacity, enter the cal weight with the keypad. When the weight
is stable press
ENTER
.
Any calibration weight is possible. MSI recommends a cal weight of at least 10% of capacity for
industrial systems. For Legal-for-trade systems MSI recommends using at least 50% of capacity,
and consider multipoint calibration.
11) “SPAN CALIBRATION 2” – Now that the CellScale has 1 span
point, it transmits back the current reading. If a single span point
is adequate, press
ESC
(
SETUP
) to end calibration. If multipoint
calibration is desired, load the next cal weight. If the displayed scale
weight agrees with the cal weight, you do not need to calibrate
the point. However if the weight displayed does not match the cal
weight, use the keypad to enter the correct weight. When the weight
is stable press
ENTER
.
12) Repeat step 11 up to a total of 9 span points. Span points can be
in any order, low to high, high to low, jumping around...whatever.
At any time, pressing the
ESC
(Setup) key will end calibration.
Even non-linear transducers can be linearized with the judicious
application of multi-point calibration.