APPENDIX D: DISPLAYED ERROR CODES

CODE

MODE

MEANING / POSSIBLE SOLUTION

Normal Operating

Mode

Gross Overload. A weight greater than the rated capacity has been

applied to the scale. Remove the weight from the platter or try re-

calibrating the scale. Otherwise, check for a bad load cell connection or

possible load cell damage due to overloading.

Err 0

Span Calibration

Mode (F17)

Keyed-in weight value is larger than full scale capacity. Use a smaller test

weight or check keyed-in value.

Err 1

Span Calibration

Mode (F17)

Keyed-in weight value is less than 1% of full scale capacity. Use a larger

test weight or check keyed-in value.

Err 2

Span Calibration

Mode (F17)

There is not enough load cell signal to produce the internal counts

necessary to properly calibrate the scale. First check all load connections.

Use F16 mode to view internal counts. See Appendix C for more

information.

Err 3

All Modes

Non-volatile memory read error. One or more setup parameters have

been lost.

Err 4

All Modes

Non-volatile memory write error. Indicator needs service.

Err 5

Key-in Span

Calibration Mode

(F20)

You have attempted to enter a zero value for C1. Enter a known

calibration value greater than zero.

Err 7

Initialization

No reading from the ADC. Make sure there is a load cell(s) connected to

the indicator at start-up.

Err 9

Normal Operating

Mode

Span calibration value has been lost. Re-calibrate the scale.

t Err 1

Normal Operating

Mode

You have attempted to key-in a tare weight that is greater than the scale’s

maximum capacity. For example, on a 6,000 lb scale, you cannot enter,

say, 7,000 lb.

t Err 6

Normal Operating

Mode

You have attempted to key-in a tare weight whose resolution is higher

than the resolution of the scale For example, on a 100 x 0.02 lb scale, you

cannot enter, say, 10.01 lb since the resolution is 0.02 lb.

Err 10

Initialization or Setting

Code Entry

Indicator cannot read serial number. Indicator must be returned for

service.

Err 11

Initialization

Serial number mismatch. Re-enter setting code from product ID tag.

Err 12

Initialization or Setting

Code Entry

Invalid setting code. Re-enter proper setting code from product ID tag.

Err 13

Setting Code Entry

Non-volatile memory is in write-protect mode. Toggle position of

Setup/Calibration Switch or JP2 jumper (See Chapter 3).

Note:

Contact Service Department at Salter Brecknell for instructions on how to re-enter setting

code.