NAME/CODE

DESCRIPTION

CODE/VALUE

F9

Display

Divisions

Determines the desired weight increments. Value should be

consistent with legal requirements.

1

√√√

2

5

F10

Decimal Pt.

Determines location of the decimal point.

0

√√√

0.0

0.00 0.000

0.0000 00

F11

Custom Unit

Multiplier

Actuates the function that allows you to enter a multiplier for the

custom unit. Scrolling down with the

#2

key one level begins the

procedure.

Press

#2

key to

begin sequence

1

√√√

F12

Custom Unit

Enable

Used to enable or disable the custom unit whenever the UNITS key is

pressed. This feature must be disabled for commercial applications.

"0" = Disable custom unit "1" = Enable custom unit

0

1

√√√

F13

Piece Count

Enable

Used to enable or disable the piece count feature. This feature must

be disabled for commercial applications.

"0" = Disable piece count "1" = Enable piece count

0

1

√√√

F14

lb/kg

Conversion

Enable

Used to enable or disable the lb to kg conversion whenever the

UNITS key is pressed.

"0" = Disable conversion "1" = Enable conversion

0

1

√√√

F16

Zero

Calibration

Places indicator into the zero calibration routine. Scrolling down with

the

#2

key one level begins the procedure.

Press

#2

key to

begin sequence

F17

Span

Calibration

Places indicator into the span calibration routine. Scrolling down with

the

#2

key one level begins the procedure.

Press

#2

key to

begin sequence

F18

View Calibra-

tion

Actuates the function that allows you to view both the zero and span

calibration value. The values displayed in this function are valid only

after Calibration (F16 & F17) has been successfully completed.

Scrolling down with the

#2

key one level begins the procedure.

Press

#2

key to

begin sequence

F19

Key-in Zero

Allows you to key-in known zero calibration value in case of memory

loss in the field. Scrolling down with the

#2

key one level begins the

procedure.

Press

#2

key to

begin sequence

F20

Key-in Span

Allows you to key-in a known span calibration value in case of mem-

ory loss in the field. Scrolling down with the

#2

key one level begins

the procedure.

Press

#2

key to

begin sequence

F21

Factory Reset

This sub-menu will reset all parameters in the “F” and “A” menu to the

default settings. USE WITH CAUTION!

Press the

#2

key

twice to execute.