6. If the calibration was

not

successful, one of the error messages below will appear. Take

the indicated action to correct the problem, then perform a new calibration.
"

Err0

" - The calibration test weight or the keyed-in weight is larger than the full capacity

of the scale. Change the calibration test weight or check the input data.
"

Err1

" - The calibration test weight or the keyed-in weight is smaller than 1% of the full

capacity of the scale. Change the calibration test weight or check the input data.
"

Err2

" - The internal resolution of the scale is not high enough to accept the calibration

value. Select a larger parameter for the Span Gain (F2). SEE APPENDIX C FOR

MORE INFORMATION.

6.4

VIEW CALIBRATION VALUES (F18)

Note:

The values displayed in this procedure are valid only after a successful calibration

has been performed using F16 and F17.

1. While in the Setup mode, scroll to "

F 18

", then scroll down once using the #2 key to

enter View calibration menu.

2. The display will momentarily show "

CAL 0

" followed by a value. This value is the

zero

calibration value

and should be recorded in the table below. Press any key to

continue.

3. The display will momentarily show "

CAL 1

" followed by another value. This value is the

span calibration value

and should also be recorded in the table below. Press any key

to return to upper level (F18).

INDICATOR

ZERO CALIBRATION VALUE

SPAN CALIBRATION VALUE

S/N:

Table 6-1: Calibration Value Table

6.5

KEY-IN ZERO CALIBRATION VALUE (F19)

Note:

This procedure is intended for emergency use only in the case of non-volatile

memory loss. A valid zero calibration value, obtained from a successful F16

calibration procedure, must be used.

1. While in the Setup mode, scroll to "

F 19

", then scroll down once using the #2 key. The

display will momentarily show "

ET

C 0

", followed by a value of zero

2. Use the front panel keys to key-in the actual zero calibration value. If you make a

mistake, press the CL key to clear your entry and start over.

3. After entering the exact value, press the NET/GROSS key to save the value. The

display will show "

E C 0

" momentarily, then revert back up to F19.