Salter Brecknell 800 Series User Manual
Page 26

6. If the calibration was
not
successful, one of the error messages below will appear. Take
the indicated action to correct the problem, then perform a new calibration.
"
Err0
" - The calibration test weight or the keyed-in weight is larger than the full capacity
of the scale. Change the calibration test weight or check the input data.
"
Err1
" - The calibration test weight or the keyed-in weight is smaller than 1% of the full
capacity of the scale. Change the calibration test weight or check the input data.
"
Err2
" - The internal resolution of the scale is not high enough to accept the calibration
value. Select a larger parameter for the Span Gain (F2). SEE APPENDIX C FOR
MORE INFORMATION.
6.4
VIEW CALIBRATION VALUES (F18)
Note:
The values displayed in this procedure are valid only after a successful calibration
has been performed using F16 and F17.
1. While in the Setup mode, scroll to "
F 18
", then scroll down once using the #2 key to
enter View calibration menu.
2. The display will momentarily show "
CAL 0
" followed by a value. This value is the
zero
calibration value
and should be recorded in the table below. Press any key to
continue.
3. The display will momentarily show "
CAL 1
" followed by another value. This value is the
span calibration value
and should also be recorded in the table below. Press any key
to return to upper level (F18).
INDICATOR
ZERO CALIBRATION VALUE
SPAN CALIBRATION VALUE
S/N:
Table 6-1: Calibration Value Table
6.5
KEY-IN ZERO CALIBRATION VALUE (F19)
Note:
This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid zero calibration value, obtained from a successful F16
calibration procedure, must be used.
1. While in the Setup mode, scroll to "
F 19
", then scroll down once using the #2 key. The
display will momentarily show "
ET
C 0
", followed by a value of zero
2. Use the front panel keys to key-in the actual zero calibration value. If you make a
mistake, press the CL key to clear your entry and start over.
3. After entering the exact value, press the NET/GROSS key to save the value. The
display will show "
E C 0
" momentarily, then revert back up to F19.