beautypg.com

Experion automated electrophoresis system – Bio-Rad Experion DNA Analysis Kits User Manual

Page 46

background image

42

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Error

Probable Cause

Recommended Action

Peaks are small, broad, or
missing in some samples,
but present in other samples
on the chip

Contaminants are present

Prepare samples in a different buffer or dilute them in water,
if necessary

Do not use autoclaved water for diluting samples or
ladder. Use only high-quality 0.2 µm-filtered water (such as
Experion DEPC-treated water)

Late migration (peaks
broaden and are delayed
over the course of the run)

Turn analysis off (see Section 6.7, Turning Analysis Off, for
more information). Late migration is indicated if there is a
drift in the migration of the lower marker across the chip and
the upper marker migrates later and runs off the top of the
gel view. Refer to the troubleshooting tips that follow

Stop the run, remove the chip, and use a clean pipet tip to
remove or dislodge the bubbles, or remove and replace the
solution in affected wells

When pipetting, insert the tip vertically and to the bottom of
the well. Dispense liquids slowly. Do not expel air at the end
of the pipetting step. Dispense only to the first stop

Air bubbles are interfering with electrical
contact in one or more of the wells

Peaks migrating faster than
usual; electropherogram
appears compressed

Electrophoresis station temperature is
inappropriate or fluctuating during the run
(should be 30–35°C); there is no cooling
unit in the electrophoresis station, so if the
temperature changes during the course of a
run, the samples exhibit different separation
characteristics

Ensure the temperature of the room is appropriate and
stable, and place the electrophoresis station away from all
heat sources, such as windows or ovens

Sizing is incorrect

DNA ladder peaks and upper and lower
markers were improperly assigned by the
software

Exclude peaks (follow the instructions in Section 6.3,
Excluding a Peak From Analysis) or manually assign m
arkers
(follow the instructions in Section 6.2, Manually Setting a
Marker), if necessary

Quantitation is incorrect

Upper and lower markers were not properly
assigned by the software

Check that both markers were properly assigned. Manually
select the marker(s), if necessary (follow the instructions in
Section 6.2, Manually Setting a Marker)

Pipetting and/or dilution errors occurred during
sample preparation or chip loading

Ensure calculations and dilutions are correct

Ensure pipets are calibrated, and use pipets that accurately
deliver volumes of ≤10 µl

Do not modify the sample preparation and loading protocols
described in this manual

Ensure that the total DNA concentration of samples is within
the linear dynamic range before proceeding with the sample
preparation instructions described in Section 3.4; if total
DNA concentration is unknown, run a DNA analysis of a
dilution series of the sample

Review the essential practices described in Chapter 2
before initiating another analysis with a new chip

Samples are old or have not been stored
properly, and the DNA degraded

Prepare fresh samples and try the analysis again

DNA is <100 bp

Small DNA fragments may bind dye differently, depending
on base composition

Ghost peaks or
contaminants appear in
electropherograms

Electrodes are contaminated

Follow the deep cleaning procedure described in Appendix
B and the software Help section (search term “electrodes”)

Use only 0.2 µm-filtered water (such as Experion DEPC-
treated water) for diluting DNA samples and the DNA ladder;
do not use autoclaved water

Experion Automated Electrophoresis System

This manual is related to the following products: