beautypg.com

8 manual peak integration – Bio-Rad Experion DNA Analysis Kits User Manual

Page 42

background image

38

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Fig . 6 .4 . Manual integration mode indicates the beginning and end of peak baselines with blue dots .

6.8 Manual Peak Integration

This function enables definition or refinement of peak areas for the ladder and all sample lanes.

1. Select a sample and click or select Analysis > Manual Integration to start manual integration.

2. Adjust the baseline by clicking and dragging the dots at the ends of the peaks (Figure 6.4).

If a peak in the electropherogram is not detected by the software, right-click on that peak and select
Add Peak. Zoom in on the peak for a better view of the baseline.

3. Save any changes by selecting File > Save Run As (use a new file name), or revert to the original

baseline and integration by clicking .

Manual integration applies to individual sample wells. Moving back to automated integration loses
changes made during manual integration unless those changes are saved.

Experion Automated Electrophoresis System

This manual is related to the following products: