beautypg.com

Experion dna 1k and dna 12k analysis kits – Bio-Rad Experion DNA Analysis Kits User Manual

Page 45

background image

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

41

Error

Probable Cause

Recommended Action

No peaks detected in
some samples

Samples were not prepared properly

Review the sample preparation procedure detailed in
Chapter 3

DNA stain was photobleached

Protect the DNA stain and gel-stain solution (GS) from light

Prepare new GS; use a new tube of DNA stain if necessary

Particulates are clogging the microchannels

Use only high-quality, 0.2 µm-filtered water (such as
Experion DEPC-treated water, not autoclaved water)

Verify that the gel (G) and GS were properly filtered

If using samples that contain particulates, perform a quick
microcentrifuge spin of the prepared sample to pellet
particulates before loading

No peaks detected in any
samples

Gel-stain solution (GS) is old or has been
photobleached

Prepare fresh GS

Chip was not primed with the correct pressure
and time settings

Check settings on the priming station and repeat the
analysis with a new chip

Particulates are clogging the microchannels

Use only high-quality 0.2 µm-filtered water (such as
Experion DEPC-treated water, not autoclaved water)

Verify that the GS was filtered properly

If using samples that contain particulates, perform a quick
microcentrifuge spin of the prepared sample to pellet
particulates before loading

Lower and/or upper marker
is missing

Samples do not contain sample buffer or were
not prepared properly (if both are missing)

Review the sample preparation procedure in Chapter 3

Late migration

If only the upper marker is missing, refer to the
troubleshooting tips for late migration (below)

Electropherogram peaks are
much smaller than expected
(peaks are <50% of what is
expected)

Too much stain in the gel-stain solution (GS);
sample cannot destain completely, and the
background is high; too much stain can
contribute to high background and low sample
and ladder peaks

Prepare new GS using a new tube of stain. Ensure stain is
thawed completely before preparation

Not enough sample was added

Ensure pipets are calibrated

Use pipets that accurately deliver volumes of ≤10 µl

Too much salt in the sample; salt ions compete
with the sample ions during injection

Analyze a dilution series of the sample in water on another
chip. If the peak heights and calculated DNA concentrations
of the diluted sample increase, the salt is too high in the
original sample. Determine the proper dilution to minimize
the salt effect

Gel-stain solution (GS) is old or has been
photobleached

Prepare fresh GS

Reagents, specifically the DNA stain, were not
brought to room temperature prior to use; if
stain is not equilibrated to room temperature, it
will be more concentrated when it is added to
the gel, which may result in significantly lower
peaks and areas

Repeat the analysis with a new chip, new tube of stain, and
properly equilibrated reagents

Gel-stain solution (GS) was mislabeled or
used improperly. DNA chips contain specific
wells for GS that carry the sample through
microchannels and supply dye for analysis

Ensure chip was primed with GS

Ensure GS was used in all wells marked GS

Experion DNA 1K and DNA 12K Analysis Kits

This manual is related to the following products: