6.7 Turning Analysis Off

To turn analysis off, select Analysis > Turn Analysis Off. Analysis remains off for all runs until it is
turned back on.

Turning analysis off displays the raw data from a run; it deactivates alignment, shifting the positions of
the upper and lower markers in each sample. In addition, the pink arrows (in the virtual gel) indicating
the positions of the markers appear only in the ladder lane. With analysis off, the migration of the
markers may not be perfectly aligned and the position of the upper marker may drift (Figure 6.3). Turning
analysis off can help reveal if a problem is isolated (one or a few wells have markers out of alignment) or
is general (alignment problems across the entire chip).

Fig . 6 .2 . Algorithm Inclusion Settings options .

Fig . 6 .3 . Effects of turning analysis off .

Analysis on

Analysis off