1.5 Safety

Caution/Warning

The PROTEAN IEF Cell uses high output voltages that are electrically isolated from
earth ground to minimize the risk of electrical shock to the user. The following
guidelines should be observed and followed when using the power supply.

The PROTEAN IEF Cell has passed tests for operation at temperatures between 10 °C
and 35 °C, with relative humidity between 0 and 90% non-condensing. Operating the cell
outside these conditions is not recommended by Bio-Rad and will void the warranty.

This instrument is intended for laboratory use only.

1. To insure adequate cooling of the cell’s power supply, be sure that there is at least 6 cm

clearance around the unit. Do not block the fan vents at the rear or underneath the unit.

Warning!
Do not use cloth or absorbent pads under the unit. These or other loose items may be pulled
into the fan intake, causing damage to the unit due to overheating, and voiding the warranty.

2. Always connect the cell to a 3-prong, grounded AC outlet, using the 3-prong AC power

cord provided with the cell.

3. Do not operate the cell in extreme humidity (>90%) or where condensation can short the

internal electrical circuits of the cell.

4. Disconnect power to the PROTEAN IEF Cell before servicing. No user-serviceable parts

are inside the instrument. Contact Bio-Rad service personnel for service.

5. The PROTEAN IEF Cell is certified to meet the EN61010-1 safety standard for safety of

laboratory equipment. Certified products are safe to use when operated in accordance
with the instruction manual. This safety certification does not extend to other equipment
or accessories not EN61010-1 certified, even when connected to the PROTEAN IEF Cell.

6. The PROTEAN IEF Cell conforms to the class A standards for Electromagnetic

Emissions, intended for laboratory equipment applications. It is possible that emissions
from this product may interfere with some sensitive appliances when placed nearby or on
the same circuit as those appliances. The user should be aware of this potential and take
appropriate measures to avoid interference.

7.

This instrument should not be modified or altered in any way. Alteration of this instrument
will void the manufacturer’s warranty, void the EN61010-1 certification, and create a
potential safety hazard for the user.

8. Bio-Rad is not responsible for any injury or damage caused by the use of this instrument

for purposes other than those for which it is intended or by modifications of the instrument
not performed by Bio-Rad or an authorized agent.

EN61010-1 is an internationally accepted electrical safety standard for laboratory instruments.